The high resolution user-friendly screen displays real-time flow/volume loops and detailed measurement matrices.The data can be either printed within seconds or transmitted to a PC.
Disposable Sensor SP-150
With SCHILLER’s unique SP-150 disposable sensors, the risk of cross-contamination is reduced to a minimum. The small, lightweight sensors are low-cost and completely disposable. Simply remove, discard and replace the sensor after each patient performance. No need for desinfection, easy handling and saving time: these are maximum benefits for both patient and physician.
Reusable Sensor SP-20
SCHILLER’s SP-20 reusable sensors are small, lightweight and very easy to clean. The low-cost filters provide the most economical method of performing pulmonary function tests.
Software Options:
– SEMA PC software for validation and archiving of test data on PC
– Memory for up to 100 tests
Measured Values:
FVC: FVC, FEV0.5, FEV1.0, FEV3.0, FEV0.5/FVC, FEV1.0/FVC, FEV3.0/FVC, FEF0.2–1.2,
FEF25–75%, FEF75–85%, PEF, MEF75%, MEF50%, MEF25%, FIVC, FIV1.0, FIV1.0/FIVC,
FIV1.0/FVC, PIF, MIF50%
SVC: SVC, ERV, IRV, TV
MV: MV, RR, TV
MVV: MVV, RR, TV
Prediction Equation:
– Adults: ECCS, Crapo, Morris, Knudson, Knudson 76, Austria, Berglund, Finnish,
Indian, Composite, Polgar
– Children: Quanier&Tammeling, Austria, India, Knudson, Knudson 76, Polgar
– Comparison pre/post medication possible
– Extrapolated predicted values
Presentation:
– Flow/volume loop
– Flow/time plot
– Measurement values table
Standards Compliance: ATS 94, OSHA, NIOSH
SPIROVIT SP-150 Pneumotach Flowsensor for Pulmonary Function Testing with
Disposable Sensors:
Dimensions: 118 x 36 x 28 mm, app. 120 g; 4.6 x 1.4 x 1.1 in., app. 0.26 Ibs
Measuring Method: Pneumotachometer
Measuring Ranges:
– Flow: 0 to ±16 l/s
– Volume: 0 to ±15 l
Measuring Accuracy: According ATS 1994
Flow Impedance: <0.2 mbar *s/l at 12 l/s
SPIROVIT SP-20 Pneumotach Flowsensor for Pulmonary Function Testing with Reusable Sensor:
Dimensions: 125 x 36 x 28mm, app. 160g; 4.9 x 1.4 x 1.1 in., app. 0.34 Ibs
Measuring Method, Measuring Ranges, Measuring Accuracy, Flow Impedance
same as SP-150
All tech
– 1 pneumotach sensor (choice of SP-150 disposable or SP-20 reusable sensor)
– 1 noseclip
– 1 pack of disposable mouthpieces (for SP-150) or filters (for SP-20)
– 1 power cable
– 1 pack of paper
– 1 operating manual
Hardware Options: Transport case, calibration syringe (required)